Knowledge Base · Fault codes & symptoms

BW500 error 221: the span the scale would not accept

BW500 error 221 means the integrator refused your span calibration because it moved too far from the stored reference. What causes it, and the fix.

The short answer

Error 221 is a Milltronics BW500 refusing a span calibration, because the new figure sits too far from its stored reference span, around 12.5%. Nothing in the integrator has failed. Load on the belt no longer gives the signal the scale was taught, so find out what changed before anyone forces a new span through.

This guide covers what 221 compares, how to rule out the test load, what moves a span that far, and how to settle which span is right. For the rest of the BW500’s codes, see our BW500 fault codes and calibration guide. For bookings and parts, see Siemens Milltronics belt scale and BW500 servicing. We’re an independent service company, not a Siemens dealer or agent.

What is error 221 actually comparing?

Your new span against a stored initial span, counted cumulatively. The deviation shown after a span run is the movement since the previous span, and it’s usually small. The limit that trips 221 is the total drift back to the initial span: the reference set at commissioning, or the last time someone declared a new one.

So a scale can pass span after span, moving a little each time, then refuse on the day the tally crosses the limit. On a multispan unit, each span has its own reference.

Is 12.5% a lot?

It’s a huge amount of tonnage. Span error is proportional, so it grows with throughput. On a conveyor running 2,000 t/h, 12.5% is 250 t/h. The manual suggests investigating a deviation from around 3%, so 221 is the last line of defence, not the first warning.

The numbers we work to: repeat span runs within about 0.1%, and a routine span that has moved more than about 0.25% needs attention. More in zero, span and repeatability explained.

Who adjusts what. On our service visits we adjust the zero and span when the checks show it’s needed. Between visits, site should run a zero about every two weeks and after a big change in the weather, and check the span but record it rather than accepting it. We recommend a service calibration every quarter so the scale keeps up with the seasons.

Was it the test load, or the scale?

Work on or near a conveyor, or inside an electrical enclosure, needs isolation under your site’s procedures and trained, authorised people. If that’s not you, stop and call us on 1300 101 666.

Start with the test load. It’s the cheapest to rule out, and it’s wrong more often than the scale is. The usual ones:

  • weights not actually bearing on the weigh frame
  • the wrong weight value entered for the reference fitted, which gives a clean, repeatable, completely wrong answer
  • a chain doubled up or laid crooked
  • weights left on from an earlier step, so the zero under the span is already wrong
  • a span run without a fresh zero, or with a prefeed still trickling material on.

Where the weights themselves are unknown, the BW500’s factoring routine can value them against the current span. See test weights, chains and billets.

What kind of change moves a span that far?

A change in how load reaches the load cells, or how their signal reaches the integrator. Roughly in order:

  • A load cell replaced, or one quietly failing. Fitted without rebalancing and recalibrating, a new cell is a classic 221 at the next service. See is it the load cell, the cable or the junction box?
  • Belt work. A new belt, splice or retension changes mass per metre, stiffness and sag, so load reaches the frame differently. After a new belt or splice, re-measure belt speed and length, re-enter them and recalibrate. See why span changes after belt work.
  • Geometry. A frame knocked in a shutdown, idlers moved or swapped for a different type, or tie rods corroded stiff so the frame can’t float. Weigh idlers should typically sit within about 0.3 mm of the idlers either side; the exact target depends on the scale.
  • Wiring. Water in the junction box, a chafed cable, or sense leads not used on a long run, so voltage lost in the cable reads as a moved span.
  • Setup data re-entered wrong. Belt length, speed, idler spacing and design load all feed the calculation, so a typo after an integrator swap moves the span.

Why is declaring a fresh initial span the wrong move?

It overwrites the reference every later calibration is judged against, so the alarm goes quiet because the fault can no longer be seen. The scale’s answer to “how far have you moved?” resets to nothing from wherever it is today, faults included. The manual says to use it sparingly, and only after a thorough mechanical check. It’s the last step, not the first, and it’s the most common way a site loses its calibration history.

Replacing the integrator or reloading its software is different. The BW500 has a direct entry route for exactly that, so recorded counts go back in and the reference survives the hardware change. That only works if somebody wrote the counts down.

How do you settle whether the new span or the old one is right?

With material. A material test runs a known tonnage over the scale and checks the total against an independent weight, on a decent sample: at least half design rate for several minutes. On a BW500 the correction goes in as a percentage adjustment and the unit reports a revised test load value, so your on-site reference comes back into line without touching the initial span.

Run it twice. Two tests that agree justify the adjustment. Two that disagree say the scale isn’t stable, and calibrating an unstable scale locks in a bad number. See belt weigher calibration methods.

What are your tonnes doing meanwhile?

They’re still being counted on the old span. A refused calibration changes nothing, so the integrator keeps totalising with the figure it already had. That’s safe but not neutral: the span you just measured is evidence the stored one is wrong. Note the date it appeared and treat that conveyor’s totals as provisional until it’s settled.

Need a manual for your equipment?

Check whether the nameplate says BW500 or BW500/L. The /L has no multispan, so a full-unit manual will send you looking for per-span references your unit doesn’t have. A manual badged Milltronics and a nameplate reading Siemens are usually the same unit.

Talk to us about manuals for your equipment. Send the make, model, serial number and software version (a photo of the label is fine) and we’ll help you find the right document.

When should you call us?

Call when 221 comes back after the test load has been ruled out, when a load cell has been changed or belt work done, when nobody can find the recorded counts, when repeat tests won’t agree, or when someone is under pressure to declare a fresh initial span before the cause is known. That last one is worth a call the same day: 1300 101 666. See Siemens Milltronics belt scale service.

Frequently asked questions

Can I just force the span through on a BW500?

You can, by declaring a new initial span, but don’t. It re-bases the reference on today’s reading, so a real fault stops being visible and every later calibration is judged against a bad figure.

Does error 221 stop the BW500 totalising?

No. The refused span is thrown away and the integrator carries on with the figure it had. The plant keeps running, but treat the tonnes as suspect.

How is 221 different from a span that’s too small to use?

A span too small to use means there isn’t enough signal at design load to work with, usually because the reference isn’t on the scale or the installation figures are wrong. 221 means the signal is there but disagrees with the stored reference.

This guide is general information based on our field experience. Every installation is different, so check your equipment’s documentation and your site procedures before acting on it, and talk to us if you’re unsure.

Reviewed by the Accurate Industries service team. Last updated 8 November 2026.