The BW500 is the Milltronics integrator that turns a belt weigher’s load and speed signals into a rate in t/h and a running total. Almost every code it shows is reporting something outside the box: a load cell pair that disagrees, a speed signal that doesn’t match the setup, or a span that moved because the scale under the belt moved first. So start at the belt, not the box.
This guide covers the codes we see most, what a BW500 calibration involves, and the speed, load cell and comms points that catch people out. For bookings, parts and older Milltronics gear, see Siemens Milltronics belt scale and BW500 servicing. We’re an independent service company, not a Siemens dealer or agent.
What is the BW500, and which one have you got?
It’s Milltronics’ full-featured belt scale integrator. The cut-down BW500/L has fewer relays, one analogue output, two load cell inputs instead of four, and no batching, PID or multispan. Check the nameplate, because they don’t share a manual.
Both sit in an IP65 field enclosure rated to a 50 °C ambient, which matters on a Pilbara or Bowen Basin conveyor in full sun. It’s a long-lived design, so a BW500 on one conveyor and much older Milltronics gear on the next is normal.
What do the common BW500 error codes mean?
Write down the code, the message and what happened just before it. Then work through it like this:
| Message | What we check first | When to call us |
|---|---|---|
| 201 or 202: load cell error | Water in the junction box, a chafed cable, a loose terminal. Then compare each cell’s raw millivolts. 201 is the A and B pair, 202 the C and D pair. | If it comes back once the box is dry and resealed. See BW500 error 201 and 202. |
| Span signal too small to use | Were the weights or chain actually on the scale? Do the setup figures match the installation? On a new scale, are the transit brackets off? | If it stays low with the reference correctly applied. |
| 221: span out of range | The new span is too far from the stored reference (around 12.5%). Rule out the test load, then look for a changed load cell, belt work, geometry or wiring. | Before anyone declares a fresh initial span. See BW500 error 221. |
| 222: auto zero out of range | Build-up, carryback, a stiff tie rod or a dragging idler. Clean and inspect before re-zeroing. | If it comes straight back after a proper clean. See BW500 error 222. |
| 212 or 213: maximum speed or rate exceeded | Time a marked length of belt against the display. Check the speed constant against the sensor and pulley actually fitted. | If the timed speed disagrees with the display, or the plant now runs above design rate. See BW500 error 212 and 213. |
| Access level (security) message | Someone tried a change the current access level doesn’t allow. Nothing is broken. Some units also have an internal switch that locks calibration values. | If nobody on site has the access you need. See locked out of your integrator? |
| Low battery or power failure flag | The memory backup battery is going, or the supply was interrupted. Calibration sits in non-volatile memory; the battery protects the totals and clock. | Book a battery in. Call if power failures keep recurring. |
Resetting a code over and over hides the evidence without fixing anything. If the same one keeps coming back, the cause never went away.
What does a BW500 calibration involve?
Work on or near a conveyor, or inside an electrical enclosure, needs isolation under your site’s procedures and trained, authorised people. If that’s not you, stop and call us on 1300 101 666.
- Record the as-found condition: settings, recent zero and span figures and any codes.
- Inspect the scale. Weigh idlers in line with the idlers either side (typically within about 0.3 mm; the exact target depends on the scale), the frame free, the belt tracking centrally, sensors and wiring sound.
- Check belt speed and length. More than about 2% between a timed speed and the display needs investigating. After a new belt or splice, re-measure both and re-enter them.
- Zero on a clean, empty belt after a few minutes running at normal speed. The BW500 averages over whole belt revolutions, so the belt length settings have to be right.
- Span against a known reference: test weights, a calibration chain, or the unit’s electronic calculation from load cell data, idler spacing and incline. Only weights and chain put real force through the frame. See belt weigher calibration methods.
- Prove it with material. Two material tests that agree justify adjusting the span. Two that disagree mean the scale isn’t stable, and adjusting then locks in a bad number.
- Record the as-left figures and anything still limiting the result.
The numbers we work to: repeat span runs within about 0.1%, and a routine span that has moved more than about 0.25% needs attention. The BW500’s own 221 limit, around 12.5%, is a last line of defence, not a tolerance.
Who adjusts what. On our service visits we adjust the zero and span when the checks show it’s needed. Between visits, site should run a zero about every two weeks and after a big change in the weather, and check the span but record it rather than accepting it. We recommend a service calibration every quarter so the scale keeps up with the seasons. See what happens during a calibration visit.
Why does the zero or span move on a BW500?
Because something under the belt changed. Zero moves with build-up and carryback, a binding tie rod, a seized idler, belt tracking or water in a junction box. After a proper clean, a one-off zero change of around 1% is normal. A zero that keeps walking needs investigating, not re-zeroing. A new belt takes a week or two to bed in, so check the zero after the belt work and again once it has settled. See why a belt weigher zero drifts.
Span moves when the load-to-signal relationship changes: a load cell, frame geometry, belt tension or the setup data. Zero and span moving together points at geometry rather than dirt. More in zero, span and repeatability explained and common belt weigher faults. One BW500 point: the full unit holds up to eight spans, so totals that are right on one material and out on another may need a second span, not a fault hunt.
What speed sensor and load cell points matter?
Speed. The speed input has a hardware switch for two logic levels, roughly 5 V or 10 to 24 V. Sensor type, polarity and switch position all have to agree; the NPN versus PNP trap is in our belt weigher speed sensors guide. The unit shows the incoming pulse frequency in hertz, so you can see whether pulses are arriving before you pull anything apart. Where there’s no speed sensor the terminals are jumpered, and a jumper left in keeps the unit totalising with the belt stopped.
If the sensor needs replacing, our own USR speed sensor mounts on the tail drum or a trailing arm, and its multiple outputs let one sensor feed two belt weighers, or a belt weigher plus the conveyor underspeed. We’ll confirm it suits the BW500 input first.
Load cells. The BW500 excites the cells at 10 V and shows the raw millivolts per cell. Long runs need the sensing (six- or eight-wire) arrangement, or voltage lost down the cable reads as a calibration error that wanders with temperature. Balance the cells electronically before the first calibration and whenever one is refitted or replaced, then zero and span again. Skipping that is a common reason a scale that was fine comes back wrong. See our load cells guide.
How does the BW500 talk to the plant?
Through a 0/4 to 20 mA output for rate, load or speed; relays for alarms and control (five on the full unit, two on the /L); remote totaliser contacts that pulse the PLC; and three serial ports (two RS-232, one isolated RS-485) speaking Modbus RTU or ASCII. Modbus TCP, EtherNet/IP and Profinet need a plug-in module.
The failures are the ordinary serial ones: settings that don’t match at both ends, comms cable tied to power cable, shields earthed at both ends. One BW500 habit: port changes don’t take effect until the unit is powered down and back up. See getting belt weigher data into your PLC or SCADA.
Need a manual for your equipment?
Check the nameplate first: the BW500 and the BW500/L share a name, not a manual. Note the model and the software version shown at power-up.
Talk to us about manuals for your equipment. Send the make, model, serial number and software version (a photo of the label is fine) and we’ll help you find the right document.
When should you call us rather than keep resetting it?
Call when the same code comes back after you’ve dealt with the obvious cause, or when clearing it would mean overwriting a calibration reference: a span refused as out of range, a zero that walks back after a proper clean, load cell errors that persist once the box is dry, a speed fault that survives a check of the sensor and wiring, a unit that has lost its configuration, or totals that no longer line up with the weighbridge. Note the code and the as-found figures first, then call us on 1300 101 666.
We supply parts for Siemens Milltronics and the other major belt weigher brands: a genuine part where it’s available, or a suitable alternative, with compatibility confirmed and the part identified on the quote. We’ll tell you straight whether to repair it or replace it. See booking a belt weigher service.
Frequently asked questions
Can a BW500 be calibrated without test weights?
Yes, with a calibration chain or the electronic calculation from load cell and installation data. Only a physical reference puts real force through the frame, and only a material test proves the whole system.
Why does my BW500 keep totalising when the belt is stopped?
Usually the speed input is jumpered, or a contact standing in for a speed sensor is stuck closed, so the unit thinks the belt is always moving. It’s quick to check.
Does a BW500 need maintenance of its own?
Very little beyond calibration. The memory backup battery is the one consumable, and we’d plan on changing it about every five years. What ages around it is the wiring, the seals and the sensors.