Errors 201 and 202 mean the BW500 has compared a pair of load cell inputs and doesn’t accept what it sees: the two sides disagree by more than a fixed amount, or one has gone silent. 201 is the A and B pair, 202 the C and D pair. On a wet site, suspect water in the junction box before the cell, and treat the tonnes as unusable while the code stands.
This guide covers what the unit compares, why it’s usually the wiring, what the raw millivolts should look like, and the order we work through it. For the rest of the BW500’s codes, see our BW500 fault codes and calibration guide. For bookings and parts, see Siemens Milltronics belt scale and BW500 servicing. We’re an independent service company, not a Siemens dealer or agent.
What is the BW500 comparing, and what makes 201 different from 202?
It compares the two inputs of one pair against each other continuously, with the cells excited at a nominal 10 V. The pair is the only difference between the codes, so the code gives you a location, not a diagnosis. The cut-down BW500/L has two load cell inputs instead of four, so it can only raise 201.
That location is worth having. The manufacturer’s cabling guidance for four- and six-cell scales is to run each pair on its own cable, so the two codes map onto two separate cable runs and a 202 on its own halves the search. Both at once points at something shared: the excitation supply, a common junction box or cable route, or the integrator.
The trip point is a fixed difference in the unit’s internal counts, not millivolts or a percentage of load, and it’s far wider than the gap between a balanced pair. So it won’t fire over a slowly drifting cell. By the time it shows, a leg of the measurement has broken, or the frame is loading one side and not the other.
Why is it usually the wiring rather than a failed load cell?
Because the wiring lives in the worst spot on the conveyor and the cell doesn’t. Water gets into a junction box through a perished lid seal, an open gland or a conduit that drains into it. It bridges the terminals and drops the insulation of the whole assembly with every cell healthy. Chafe, crush and a nicked sheath that wicks water inside do the same, and so does a wet or loose termination. More in is it the load cell, the cable or the junction box?
One trap is specific to this integrator. On short runs the excitation reference is made by links at the unit. Past roughly 150 m the sensing arrangement is used instead, so those links come out and extra conductors run to the scale. A half-finished job, or a sense wire broken later, reads as a dead input rather than a slow drift, and it turns up right after an install or a rewire.
What should the raw millivolts look like on each input?
The BW500 shows the raw, unbalanced signal for each input, and it costs nothing to look. Read for the comparison rather than a set number. A 2 mV/V cell on 10 V gives about 20 mV at full rated load, and a frame works well below that. On an empty belt the pair should sit close together and stay still.
| What you see | What it usually means |
|---|---|
| One input near zero, its partner sensible | An open circuit or a dead cell in that leg. |
| Both near zero | The excitation, the shared cable or the junction box. |
| Both present but far apart | One leg loaded differently (mechanical), or one degraded by moisture. |
| A reading that wanders while nothing moves | Moisture or a marginal connection, not the cell. |
More on testing cells in our load cells guide.
Can the mechanics push a healthy pair apart?
Yes, and a wiring inspection won’t find it. Anything sending load into one side of the frame and not the other reads as disagreement: a frame binding on the stringer, a tie rod stiffened by corrosion, a seized weigh idler.
Shock overload is the one to watch. A cell hit by a lump down the chute, or by someone jacking the frame without taking the load off the cells, can be permanently deformed and still read plausibly at rest, then part company with its partner as load comes on. A code that only appears when the plant runs hard won’t be cured by drying out a box.
What happens to your tonnes, and can you run on the remaining pair?
Treat the reading as unusable from the moment the code appears. The manual doesn’t say the unit stops totalising while a load cell error stands. The display alternates between the error and the run screen, the rate and total keep climbing, and the analogue output and remote totaliser keep feeding the plant a number the integrator itself doesn’t believe. The error could go either way: a dead input reads low, and a wet leakage path can read high. And don’t assume anyone was told: the manual doesn’t list these codes as driving the diagnostic alarm relay.
Running on the remaining pair isn’t an option. The scale is calibrated as a whole frame, and half a frame still carries material whether the integrator is listening or not. Production can keep running, but flag those tonnes and reconcile them from a weighbridge, a downstream scale or run time.
What order do we work through it?
Work on or near a conveyor, or inside an electrical enclosure, needs isolation under your site’s procedures and trained, authorised people. If that’s not you, stop and call us on 1300 101 666.
Cheapest and most likely first:
- Note which code it is and check the error log for a pattern. Does it follow rain, or heavy running?
- Read the raw signal on each input before disturbing anything.
- Open that pair’s junction box and look for water, corroded terminals and tired glands.
- Walk the cable run.
- Prove the excitation reaching the cells.
- Split the frame with check weights, cell by cell.
- Move the suspect leg to a spare input.
- Only then condemn a cell.
Calibration comes last, and it isn’t optional after a balance or a cell swap, because balancing invalidates the stored zero and span. Don’t quieten the display with a fresh initial zero or span: that overwrites the reference later calibrations are judged against.
Need a manual for your equipment?
Check the nameplate first. The full BW500 manual is written around four load cell inputs and both codes, so a BW500/L owner will go looking for a C and D pair that was never fitted. Note how many load cell inputs are actually landed in the unit, and roughly how long the cable run to the scale is.
Talk to us about manuals for your equipment. Send the make, model, serial number and software version (a photo of the label is fine) and we’ll help you find the right document.
When should you call us?
Call when the code comes back after a box has been dried out and resealed, when it only shows under load, when a cell may have taken a shock load, or when clearing it would mean overwriting a calibration. Ring 1300 101 666, or see Siemens Milltronics belt scale service.
We supply parts for Siemens Milltronics and the other major belt weigher brands: a genuine part where it’s available, or a suitable alternative, with compatibility confirmed and the part identified on the quote. If a cell or cable is damaged, we’ll tell you straight whether to repair it or replace it.
Frequently asked questions
Does error 201 mean a load cell has failed?
Not on its own. One pair of inputs disagrees badly or one has no signal, and water in a junction box, a damaged cable or a frame binding on one side will all do that with healthy cells.
Can I just clear the error and keep going?
Clearing it fixes nothing, and the tonnes aren’t trustworthy while it stands. If it comes back, book it in and reconcile the affected period from another source.
Why do I get 202 but never 201?
Because C and D is a separate pair, usually on its own cable run to its own end of the frame. It points at one cable, one set of terminations and the cells on that side.
Does the scale need recalibrating after a load cell is replaced?
Yes. Balance the cells electronically, then zero and span again. A new cell fitted without that step is a common reason a scale that was fine comes back wrong.